Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
photo-pu-7916_std

Agilent PR02 and PD03 Portable Leak Detectors

Applications

  • Semiconductor
  • Analytical instrumentation and research
  • Industry
  • Electric power

Features

  • Smart design with retractable handle while maintaining portability, at only 21 kg/46 lbs
  • Detachable control panel
  • On-demand operating interface
  • Intuitive and customizable menu
  • Small footprint, small size
  • Can be operated in any position
  • Large bright color touchscreen
  • Color graphics functionality
  • Password-protected display
  • Integrated SD memory card for recording, downloading and parameter settings
  • Voice synthesizer

Specifications Table

Leak Detector ModelASM 310
Pumping Speed
He
1.10 L/s
Rough
1.00 CFM
1.70 m3/hr
Run Up Time
< 120 seconds
Inlet Flange
KF25
Smallest Detectable Leak Rate
Helium
Vacuum Operation
< 5 x 10-12 mbar*L/s

Helium
Sniffer Mode
< 1 x 10-7 mbar*L/s
Units of Measurement
Pressure
mbar*L/s
Pa
atm
Torr

Leak Rate
mbar*L/s
Pa*m^3/s
atm*cc/s
Torr*L/s
s*ft^3/yr
Leak Rate Display Area Min
5.000 x 10-12 mbar*L/s
Inlet Pressure
max
15.0 mbar
1500 Pa
15.0 hPA
11.3 Torr
Languages
English
French
German
Chinese
Japanese
Korean
Spanish
Russian
Italian
Dimensions in. (mm.)
10.0 x 13.8 x 16.3 ( 254 x 350 x 415 )
Shopping Cart